Features
DOTE4-G Digital torque wrench tester with manual loader. Ideal for torque wrench calibration and adjustment.
Also available with Motor Drive Loader: Model DOTE4-G-MD
Motor Drive Retrofit Kit Available for DOTE4-G to upgrade standard model to motor drive version.
Pole holder and plate are equipped on all models, which helps keep wrenches in optimal testing postion.
Auto-zero function.
Up to 1000 measurement data items can be stored in the memory.
Peak/Run modes with 10 customizable hi/lo judgment patterns
Color Judgment Display makes it easy to confirm readings are within your set parameters
Global units of measure selection by keypad: cNm, NM, kgf.cm, kgf.m, lbf.in, lbf.ft
Sample quantity, maximum value (Hi), minimum value (Lo) and mean value can be displayed.
Manual and Auto Reset from 0.1~5.0 seconds
RS232C output and USB for easy connection to PC or Printer EPP16M3 (cables sold separately)
Uses AC100 to 240V +/10% 50/60Hz # BA-6, AC Adapter Included
Excel Receiver software sends collected data to a Excel®** for data management. (Provided as free download with purchase of cable) Note: Excel® is a registered trademark of the Microsoft Corporation.
Includes Calibration Certificate: ILAC MRA
Calibration Kits (sold separately) to check and adjust accuracy of tester on periodic basis. (or send to Tohnichi Service Center)
CE markings applied for use in the EU.
Display :Torque Display area indicates the torque value. Values are displayed in blue when an OK/ NG judgment result is OK, in red when it is LO/HI, and in white when no OK/NG Judgment is made.
Counter/Upper-limit :Value display Display a serial number or an upper-limit value.
Upper-/Lower-limit value Selection/Lower -limit value display :Displays an upper-/lower-limit value selection status for OK/NG Judgment or a lower-limit value.
▼ Mark display :RUN Lights up when RUN mode (Continuous display) is selected.
PEAK Lights up when PEAK mode (Maximum value display) is selected.
M.R Lights up when Auto memory/Reset is not set in the PEAK mode.
Lights up in the RUN mode.
A.R Lights up when Auto memory/Reset is set in the PEAK mode.
Unit ▼ appears on the unit (lbf·in, lbf·ft, cN·m, N·m, kgf·cm, kgf·m) selected.
Count forward Key :Moves the counter forward by one to read out measured data
Count backward Key :Moves the counter backward by one to read out measured data
Clear Key :In the PEAK mode, the peak value is reset. Or, the stored data is cleared. In the RUN mode, auto zero adjustment is performed.
Statistic processing Key :This is a key for selecting the number of samples, maximum value, minimum value or mean value.
Mode Key :This is a key for selecting the RUN mode or PEAK mode. The upper-/lower-limit value selection screen is displayed by holding down the key for 2 seconds or longer.
Memory/Reset button :Stores the measured data (measured torque, measurement date and time) and moves the counter forward by one after OK/NG Judgment of the measured data.
How to Calibrate a Torque Wrench
Calibration temperature
Calibration shall be performed at the ambient temperature in the 18 to 28°C range.
Temperature variations shall be within ±1°C.
Check the following points before starting calibration of a torque wrench
(1) Place the torque wrench tester on a stable workbench.
(2) For direct-reading type torque wrenches, read the measurement value in a direction perpendicular to the scale or dial (Parallax correction).
(3) For click-type torque wrenches (QL/SP, etc) do a warm-up operation by loading the maximum torque of each model for both directions 5 times before testing.
(4) For direct-reading type torque wrenches (CEM/DB/F type, etc), do a warm-up operation by loading the maximum torque of each model for both directions once, and make sure the pointer, or the digital display shows "0" after loading.
(5) Apply load at the center of the grip or at the marked loading point within the specified range
Precautions in Calibration
(1) For click-type torque wrenches (QL/SP, etc), load the initial torque up to 80% of the target torque, then apply the rest slowly and evenly. (take 1 - 3 seconds to apply the last 20%.) Pre-set type torque wrench (except single-purpose type such as QSP) must start at low level, then adjust to the target torque. (Do not adjust to the target torque by lowering from the higher level.)
(2) For direct-reading type torque wrenches (CEM/DB/F, etc), apply torque up to the target torque with attention not to exceed the target torque. If you exceed the target torque by mistake, please release torque down to zero and do the process from scratch.
We do undertake Servicing and repairing of Tohnichi DOTE4-G (-MD2) Torque Wrench Tester Tohnichi Range of products throughout india
Apart from supplying Tohnichi Electrical Testing products in India through dealers network, catering to a variety of customers ranging from Atomic energy, Electricity Boards, Electrical Laboratory, Shopping Complexes, Shopping Mall, Electrical Contractors, Engineering Industries, Turnkey Construction Companies, Railways, Defence Installations, Research Laboratories, Educational Institutes and various Research and Development laboratories of leading national and multinational companies.
We DFT TECH, are Tohnichi Authorized Distributors, Importers, Stockists, Dealers and Suppliers of Tohnichi DOTE4-G (-MD2) Torque Wrench Tester Products in Chennai, Assam, Madurai, Coimbatore, Nagaland, Kerala, Manipur, Mizoram, Haryana, Sikkim, VIsakhapatnam, Tirchy, Hosur, Anjaw, Tumkur, Mysore, Kolkatta, Salem, Poonamalle and etc..