Tohnichi BTGE-G Torque Gauge

Tohnichi BTGE-G Torque Gauge

Tohnichi BTGE-G Torque Gauge

Features
Digital torque gauge series for small torque range 2~200cN • m.
Digital torque gauge with built-in digital display.
Mini-size handheld torque gauge.
Flip-up display can be adjusted for optimal reading and accommodates various testing positions.
Built-in display gives user-friendliness and convenience of use.
Ideal for torque measurement for precision machinery, electronic devices, and various assembly components.
Physical stopper prevents accidental over-loading.
Up to 999 readings can be saved with computing function to display max/min/meantime values.
Coin-type lithium battery (CR2450) allows 10-hour continuous operation.
USB cable can be used for data transfer to PC.
Global design.
With CE mark, and it is for international use including the EU region.
Excel Receiver software is a data input tool for Excel® data management. It can be downloaded for free.
*Note: Excel® is a registered trademark of the Microsoft Corporation.

Detailed Descriptions of Various Functions
Continuous display (RUN mode)
When the counter is set at 000, it is in RUN mode where the displayed value increases as torque is applied, and returns to zero as it is released.

Maximum value display (PEAK mode)
2 kinds of PEAK modes are selectable as follows.
PEAK measuring mode (P-1)
When the counter is set at any value in the range of 001-999, it is in PEAK mode, where the torque value keeps the peak value after releasing the torque (Peak value can be captured only when the value exceeds the 15% of the max capacity torque).
Note * If you try to measure an object which may generate physical impact to ATGE sensor (such as rotary slip type torque screwdriver), it may cause it to show a very high value. Torque screwdriver measuring mode (P-2)

Torque screwdriver measuring mode (P-2) Set the counter to 000-999, torque value increases as torque is applied, and it holds the peak value as it clicks at the set torque.

Auto Zero function
In RUN mode, press the clear key to activate the auto zero function. (However, the torque load must be within about 20% of the maximum value).

When Err 9 is displayed
• Under no load condition, press the power switch and the clear key.
· If Err 9 message disappears, it functions properly.
· If Err 9 message remains displayed, press the reset switch, and then press the power switch and the clear key again. If the Err 9 message still remains displayed, the sensor or the circuit board may have a problem. Contact TOHNICHI to seek further assistance.

When Err 0 is displayed
Err 0 may appear if Auto Zero is activated while the torque value is still in unstable conditions.
• Under stable condition, press clear key again.
· If Err 0 message disappears, then it can be used normally.
· If Err 0 message remains displayed, press reset switch, press clear again under stable condition. If it still remains, there may be a sensor problem. Contact TOHNICHI to seek further assistance.

When rotating measuring object to measure
· Using the fixing groove, fix the torque gauge to the jig table.
· Holding the chuck holder fixed, open the three-jaw chuck and put the object to be measured in the chuck.
· Place the object to be measured in the center of three-jaw chuck, and while holding the chuck holder fixed, tighten the chuck.
· Turn on the power of the torque gauge.
· If the counter is 000, the continuous display (RUN mode) is set. If the counter is 001-999, the maximum value hold (PEAK mode) is set. It works bi-directionally.
· Before measurement, check the torque value on the display is zero. If it is not zero, press clear at no load condition to activate Auto Zero.
· Rotate the object to measure in RUN mode or PEAK mode and check the torque value on the torque gauge.

Rotating the torque gauge to measure
· Holding the chuck holder fixed, open the three-jaw chuck and put the object to be measured inside.
· Place the object to be measured in the center of three-jaw chuck, and while holding the chuck holder fixed, tighten the chuck.
· Turn on the power of the torque gauge.
· If the counter is 000, the continuous display (RUN mode) is set. If the counter is 001-999, the maximum value hold (PEAK mode) is set. Both clockwise and counterclockwise torque can be measured. It works bi-directionally.
· Before measurement, check the torque value on the display is zero. If it is not zero, press clear at no load condition to activate Auto Zero.
· Rotate the torque gauge in RUN mode or PEAK mode and check the torque value.

We do undertake Servicing and repairing of Tohnichi BTGE-G Torque Gauge Tohnichi Range of products throughout india

Apart from supplying Tohnichi Electrical Testing products in India through dealers network, catering to a variety of customers ranging from Atomic energy, Electricity Boards, Electrical Laboratory, Shopping Complexes, Shopping Mall, Electrical Contractors, Engineering Industries, Turnkey Construction Companies, Railways, Defence Installations, Research Laboratories, Educational Institutes and various Research and Development laboratories of leading national and multinational companies.

We DFT TECH, are Tohnichi Authorized Distributors, Importers, Stockists, Dealers and Suppliers of Tohnichi BTGE-G Torque Gauge Products in Chennai, Bangalore, Hyderabad, Pune, Mumbai, Delhi, Calcutta, Ahmedabad, Sricity, Ambattur, Guindy, Sriperumputhur, Oragadam, Tiruvallur, Trivandrum , Coimbatore, Tiruchy and etc..