Lutron CD-4322 CONDUCTIVITY METER, ATC, two displays

Lutron CD-4322 CONDUCTIVITY METER, ATC, two displays

Lutron CD-4322 CONDUCTIVITY METER, ATC, two displays

FEATURES
* 3 measurement range, 2,000 uS, 20 mS, 100 mS with auto range.
* Function : Conductivity, TDS.
* Separate probe, easy for operation of different measurement environment.
* Automatic temperature compensation range : 0 to 50 . ℃
* Carbon rod electrode for long life.
* Innovative feature with built-in automatic temperature compensation factor adjustable between 0 to 5.0% per . ℃
* Selecting " 0% per " of Temp. Coefficient Adjust, ℃ allows you to take uncompensated conductivity readings ( absolute conductivity measurement ).
* Microprocessor circuit assures high accuracy and provides special functions and features.
* Multi-display, show Conductivity and Temp. value at the same time.
* Data hold function for freezing the desired value.
* Records max. and min. value with recall.
* Build in temperature , measurement with default. ℃ ℉
* Manual and auto data logger, with flexible sampling time selection, can save max. 1,600 reading data.
* Power function can default to auto off or manual off.
* Build in the input socket for DC 9V power adapter.
* Use the durable, long-lasting components, including a heavy duty & compact ABS-plastic housing case.
* RS232/USB PC serial interface.
* Applications for Aquarium, Medical research, Agriculture, Fish hatcheries, Laboratory, Water conditioning, Mining industry, Schools & Colleges, Quality control... industry, quality control, school & college, water conditioning.

SPECIFICATIONS
Display :LCD size : 44 mm x 29 mm. Dual function LCD
Circuit :Custom one-chip of microprocessor LSIcircuit.
Function : Conductivity, TDS.
Ranges and Resolution :Conductivity range :2000 uS/20.00 mS/100 mS.
Accuracy :± ( 2% FS + 1 d )
Temperature Compensation :Automatic from 0 to 50 (32 - 122 ), with temperature compensation factor variable between 0 to 5.0% per C.
Conductivity Probe Structure :Carbon rod electrode for long life.
Memory Recall :Records Maximum, Minimum readings with recall.
Power off :Auto power off saves battery life, or manual off by push button.
Data Output :RS 232 PC serial interface.
Sampling Time of Data Logger : Manual :Push the data logger button once will save data one time. * Set sampling time to 0 second
Auto :1, 2, 5, 10, 30, 60, 600, 1800,3600 seconds.
Data Logger number :Max. 1,600-point Data logger.
Data Hold :Freeze the display reading.
Sampling Time :Approx. 1 second.
Data Output :RS 232/USB PC serial interface.
* Connect the optional RS232 cable UPCB-02 will get the RS232 plug.
* Connect the optional USB cable USB-01 will get the USB plug.
Power Supply :006P DC 9V battery ( Alkaline or Heavy dutytype ) or DC 9V adapter input. * AC/DC power adapter is optional
Power Current :Approx. DC 5.7 mA.
Weight :295 g/0.65 LB. * Include probe and batterry
Dimension :Main instrument :135 x 60 x 33 mm.
Conductivity Probe :Round, 22 mm Dia. x 120 mm length

Data Logger
The data logger function can save 1,600-point measuring data. The data logger procedures are following : a)Press the " REC button " once to start the Data record function and there will be a " REC " symbol on the display. b)Auto Data Logger ( Sampling time should select to 1, 2, 5, 10, 30, 60, 600, 1800 or 3600 seconds ) Press the " Logger button" once to start the Data Logger function. The " REC " symbol will flash per 1.5 second and the beeper will sound when save the data into the memory. Now the Data Logger function is executed. Manual Data Logger ( Sampling time should set to 0 second ) Press the " Logger button " once will save the data one time into the memory, at the same time the symbol " REC " will flash once and the beeper will sound.

Memory full
When execute the data logger function, if the upper display show " FULL " with flashing, it indicate the memory data already over 1,600 no. and the memory is full. c) During the Data Logger function is executed, press the " Logger button " once will stop the data logger function, the " REC " symbol will stop to flash. If press the " Logger button " once again will continuous the Data Logger function.

We do undertake Servicing and repairing of CD-4322 CONDUCTIVITY METER, ATC, two displays lutron Range of products throughout india

Apart from supplying Lutron Electrical Testing products in India through dealers network, catering to a variety of customers ranging from Atomic energy, Electricity Boards, Electrical Laboratory, Shopping Complexes, Shopping Mall, Electrical Contractors, Engineering Industries, Turnkey Construction Companies, Railways, Defence Installations, Research Laboratories, Educational Institutes and various Research and Development laboratories of leading national and multinational companies.

We DFT TECH, are Lutron Authorized Distributors, Importers, Stockists, Dealers and Suppliers of CD-4322 CONDUCTIVITY METER, ATC, two displays Products in Chennai, Bangalore, Hyderabad, Pune, Mumbai, Delhi, Calcutta, Ahmedabad, Sricity, Ambattur, Guindy, Sriperumputhur, Oragadam, Tiruvallur, Trivandrum , Coimbatore, Tiruchy and etc..